Abstract

Efforts aimed at conserving resources and containing costs have encouraged device manufacturers to seek alternative materials that can meet device performance requirements while using less material.

Keywords

Heat Seal, Peelable Film, Medical Packaging

Document

Metallocene Polyethylene Films as Alternatives to Flexible PVC for Medical Device Fabrication

The volume of plastics used to produce disposable medical devices and supplies is projected to grow to more than 1.6 billion pounds by the year 2000.1 Efforts aimed at conserving resources and containing costs have encouraged device manufacturers to seek alternative materials that can meet device performance requirements while using less material. Fabricators will look for opportunities to reduce the thickness, weight, or volume of device components without compromising the structural integrity or functionality of the device. For disposable devices and supplies, a reduction in raw materials will result in a direct reduction of waste.

Flexible polyvinyl chloride is one of the largest-volume film materials used in the manufacture of medical devices, and thus presents a significant opportunity for both raw material and waste reduction. Because PVC film provides a wide array of functional performance characteristics at a low cost, alternative materials will need to offer similar performance at comparable cost. Performance properties of PVC important to the medical device industry include its long, successful track record in medical applications; a breadth of properties achievable by compounding; the ability to be fabricated by RF welding and solvent bonding; the ability to be sterilized by autoclave, ethylene oxide (EtO), or gamma (although some yellowing can occur with 2,3gamma); a wide service temperature; durability and chemical resistance; and good breathability, elasticity, and clarity.Recent developments in metallocene single-site catalyst technology make possible precise control of molecular architecture and enable the production of polyolefin resins with very low densities and narrow molecular-weight distributions. Metallocene-catalyzed polyethylene copolymer resins (mPE) are currently being made with specific gravities in the range of 0.86–0.92 and comonomer content of 0–45%. Polyolefin plastomer resins formulated as ethylene-octene copolymers with less than 20% comonomer (produced by Dow Chemical Co., Midland, MI) have demonstrated enhanced toughness, sealability, clarity, and elasticity.

The toughness of mPE resins can allow for thinner, lighter-weight films, and the lower density of the mPE films results in a higher yield than is possible with PVC, producing more film area per pound. Very stable following sterilization by either radiation or EtO, mPE films provide good low-temperature flexibility and impact resistance, and have a low seal-initiation temperature.

The lower melt temperatures of the mPE resins (less than 110°C) make these films inappropriate for products requiring 4autoclave or high-temperature steam sterilization.This study was designed to test the suitability of films made using mPE resins as alternatives to flexible PVC films for medical device and appliance applications. Films used in this study were fabricated from ethylene-octene copolymer mPE resins with specific gravities between 0.88–0.90 and comonomer content of 12–20%.

EXPERIMENTAL PROCEDUR>

Table I. Film samples evaluated.
AbbreviationDescriptionFilm Thickness (mm)Film Density (g/cm3)
mPE film 1aMDF 7200 monolayer, embossed cast mPE film0.15 and 0.250.905
mPE film 2aXUR-52 monolayer, embossed cast mPE film0.15 and 0.250.895
PVC film 1bE30-194 medical-grade PVC film0.18, 0.20, 0.231.25
PVC film 2bE30-2152R medical-grade PVC film0.251.26
a Dow Chemical Co., Midland, MI.
b Ellay, Inc., City of Commerce, CA.

study. They included two mPE films with different densities and two medical-grade PVC films

that were designated by the manufacturer for use in medical collection or drainage bags. The mPE films used were 0.25 and 0.15 mm (10 and 6 mil) thick, and the PVC films were 0.18, 0.20, 0.23, and 0.25 mm (7, 8, 9, and 10 mil) thick. Since the films were embossed on one surface, thickness was reported as a nominal thickness per ASTM E 252.5

Standard physical properties evaluated for all films included tensile strength, elongation, modulus, tear resistance, puncture resistance, and barrier characteristics. The films were conditioned according to ASTM D 882 and then tested following the ASTM method, as shown in Tables IV.

Table II. Physical properties of mPE films.
 Test Units mPE Film 1
(mPE 1-6)
mPE Film 1
(mPE 1-10)
mPE Film 2
(mPE 2-6)
mPE Film 2
(mPE 2-10)
Physical PropertyTest
Method
SIEnglishTest
Direction
SIEnglishSIEnglishSIEnglishSIEnglish
Film thickness Film density Film nominal yieldASTM E 252 ASTM D 4321mm
g/cm3
m2/kg
mil
g/cm3
sq in./lb
-0.15
0.905
7.16
(6)
0.905
5030
0.25
0.905
4.55
10
0.905
3200
0.15
0.895
7.26
6
0.895
5100
0.25
0.895
4.31
10
0.895
3030
Yield tensile strengthASTM D 882MPapsiMD
TD
5.0
5.0
730
720
4.8
5.4
690
780
4.0
4.1
580
600
4.0
3.9
580
570
Ultimate tensile strengthASTM D 882MPapsiMD
TD
31.8
29.0
4610
4200
28.1
28.8
4080
4170
30.6
25.8
4440
3740
25.4
26.5
3690
3840
Ultimate elongationASTM D 882%%MD
TD
700
705
700
705
730
810
730
810
625
640
625
640
670
680
670
680
Toughness (energy to break)ASTM D 882J/m3in-lb/cu in.MD
TD
86.2
78.9
12500
11440
83.4
103.4
12100
14990
65.5
58.8
9500
8520
63.7
64.3
9240
9330
1% secant modulusASTM D 882MPapsiMD
TD
62.8
61.2
9100
8880
61.4
63.4
8900
9190
43.4
43.5
6290
6310
44.2
45.2
6410
6550
Elmendorf tearASTM D 1922gmsgmsMD1680
2090
1680
2090
3410
3690
3410
3690
810
1020
810
1020
1780
2020
1780
2020
Elmendorf tearASTM D 1922g/mmg/milMD
TD
11000
13700
280
348
13400
14500
341
369
5310
6700
135
170
7010
7950
178
202
Puncture resistanceASTM D 3763
modified
J/m3in-lb/cu in.-9.2133011.4165014.3207017.92600
Oxygen transmission rate Water vapor transmission rateASTM D 3985 ASTM E 96cm3/m2

g/m2
cm3/100 sq in.

g/100 sq in.
-2635

6.5
170

0.42
1550

3.5
100

0.23
3490

8.6
225

0.55
1920

4.9
124

0.32

Table III. Physical properties of PVC films.
 Test Units PVC Film 1
(PVC 1-7)
PVC Film 1
(PVC 1-8)
PVC Film 1
(PVC 1-9)
PVC Film 2
(PVC 2-10)
Physical
Property
Test
Method
SIEnglishTest
Direction
SIEnglishSIEnglishSIEnglishSIEnglish
Film thickness
Film density
Film nominal yield
ASTM E 252
ASTM D 4321
mm
g/cm3
m2/kg
mil
g/cm3
sq in./lb
-0.18
1.253
4.44
7
1.253
3120
0.20
1.253
3.94
8
1.253
2770
0.23
1.264
3.54
9
1.264
2490
0.25
1.260
3.34
10
1.260
2350
Yield tensile strengthASTM D 882MPapsiMD
TD
8.8
7.1
1270
1030
6.7
5.9
970
850
7.2
7.4
1050
1070
6.0
5.4
870
780
Ultimate tensile
strength
ASTM D 882MPapsiMD
TD
19.9
24.6
2890
3570
25.0
23.3
3630
3380
23.7
24.8
3430
3600
23.7
23.5
3430
3410
Ultimate elongationASTM D 882%%MD
TD
185
330
185
330
285
350
285
350
270
350
270
350
270
350
270
350
Toughness
(energy to break)
ASTM D 882J/m3in-lb/cu in.MD
TD
27.8
53.9
4030
7810
46.9
52.8
6805
7660
43.7
58.7
6340
8510
41.9
54.2
6080
7860
1% secant modulusASTM D 882MPapsiMD
TD
51.8
49.5
7510
7180
38.7
38.5
5610
5580
40.6
43.6
5880
6320
34.4
36.7
4990
5320
Elmendorf tearASTM D 1922gmsgmsMD2140
2230
2140
2230
1490
1740
1490
1740
1220
1760
1220
1760
1070
1030
1070
1030
Elmendorf tearASTM D 1922g/mmg/milMD
TD
12000
12600
306
319
7320
8580
186
218
5350
7720
136
196
4210
4060
107
103
Puncture resistanceASTM D 3763
modified
J/m3in-lb/cu in.-7.1710408.612507.611007.01020
Oxygen transmission
rate
Water vapor
transmission rate
ASTM D 3985
ASTM E 96
cm3/m2

g/m2
cm3/100 sq in.

g/100 sq in.
-62040  

26


1.7
  

Table IV. Film performance properties. (*Denotes that sample did not fail at maximum test load/height.)
 mPE FilmsPVC Films
Performance PropertyTest MethodTest ParameterUnitmPE 1-6mPE 1-10mPE 2-6mPE 2-10PVC 1-7PVC 1-8PVC 1-9PVC 2-10
Film thickness, ASTM E252-mm
(mil)
0.15
(6)
0.25
(10)
0.15
(6)
0.25
(10)
0.18
(7)
0.20
(8)
0.23
(9)
0.25
(10)
Airburst test-Time to burstsec1701842882691047210573
Resistance to burstingBS 7126-
101G
Load applied to failureN
lb
1410
316
>2300*
>516*
>2300*
>516*
>2300*
>516*
>2300*
>516*
>2300*
>516*
>2300*
>516*
>2300*
>516*
Resistance to impactBS 7126-
101H
Drop height to failurem
ft
>7.6*
>25*
>7.6*
>25*
>7.6*
>25*
>7.6*
>25*
2.7
9
2.5
8.3
2.7
9
3.1
10.2

The puncture-resistance test conducted on the film material is similar to ASTM D 3763, but is run at a lower impact speed. One sheet of film is held in a clamp that has a circular opening 45 mm (1.8 in.) in diameter. A 12.7-mm-diam (0.5-in.) spherical probe attached to a load cell on a moving cross-member is pushed through the film at a speed of 500 mm/min (20 in./min), and the resulting total energy per unit area required to puncture the film is recorded.

For the purposes of this study, liquid-collection bags were chosen as a typical medical device application. The bags are commonly made from two sheets of film welded together around the perimeter, with an inlet port sealed into one end; they may also include an outlet port and a means of measuring the volume of the liquid contents. For this study, bags were made using two rectangular plies of film totaling 432 cm2 (0.5 sq ft) each.

The collection bags were made using radio-frequency (RF) welding to bond the perimeter. PVC films and other polymers with high dielectric loss factors respond to RF energy and are commonly fashioned with RF welding equipment; however, mPE films and other low-loss polymers do not respond well to conventional RF welding. Successful RF welding of mPE films requires the addition of a mechanical "catalyst" to the existing RF equipment. The mPE bags for this study were produced using RF welding equipment (stabilized at 27.12 MHz) that had been modified with a reusable catalyst film by Plastics Welding Technology (Indianapolis). Hot-bar or impulse heat-seal equipment can also be used to form mPE films into devices.

The British Standards Institution (BSI) has developed a performance standard for liquid collection bags (BS 7126 and ISO 8669) that includes test methods for determining resistance to bursting (part 101, appendix G) and resistance to impact damage (part 101, appendix H).6 The method for determining resistance to bursting involves filling a bag with its rated volume of water (or 90% of its reference volume), sealing any openings, and placing the bag horizontally under a flat rigid plate on which weight is added to impose a 350-N (78.7-lb) force on the filled bag. After the load is applied for 1 minute, the bag is examined for leakage or bursting. The method for determining resistance to impact damage involves filling the bag to 50% of its reference volume and sealing it without entrapping air in it. The bag is then dropped from a height of 500 mm (19.7 in.)—so that the bottom hits a hard surface—and examined for leakage or bursting. The reference volume for the bags used in this study was 1240 ml; they were filled with 1116 ml of water for the bursting test and with 620 ml of water for the impact test.

The burst strength of the finished bags was tested by filling them with air at a controlled pressure and flow rate and measuring the time required to fill each bag to the point of failure. Bags fabricated with two 432-cm2 plies of film were inflated with air at a rate of 5.36 L/min and a line pressure of 0.083 MPa (12 psi) until they burst.

RESULTS
The results of the physical property tests are shown in Table II for the mPE films and in Table III for the PVC films. For the physical properties tested, the mPE films demonstrated similar or superior results when compared with the PVC film of the same thickness. Significant differences—when the mPE films showed property improvements of greater than 50% over the PVC film—were seen in results of the elongation, toughness, tear-strength, and puncture-resistance tests.

The mPE films also showed physical property advantages when compared with thicker PVC films. The properties of the 0.15-mm-thick mPE films (of two different densities—Types 1 and 2—as shown in Table I) were comparable or superior to those of PVC films of all tested thicknesses—including the 0.25-mm-thick film—demonstrating significant improvements in tensile strength, elongation, toughness, and puncture resistance.

The water vapor and oxygen transmission rates of mPE films are inversely related to their density and thickness: as the density or thickness increases, the transmission rates decrease. The barrier properties of PVC films are a function of the formulation and plasticizer content: in general, the more highly plasticized films have higher transmission rates. The mPE films tested had about five times lower water vapor transmission rates and about four times higher oxygen transmission rates than the PVC films. Normalized for thickness, the water-vapor transmission rates (gm?mm/m2/24 hr at 37°C and 90% RH) of Types 1 and 2 mPE film were 0.93 and 1.26, respectively, compared with 6.0 for Type 1 PVC film. The oxygen transmission rates (cm2?mm/m2/24 hr at 23°C and 1 atm) for the Types 1 and 2 mPE film were 390 and 500, respectively, compared with 110 for Type 1 PVC film.

The performance property tests conducted using the water-filled bags demonstrated significant differences between the mPE films and PVC film. Results are shown in Table IV and Figures 1 and 2. The resistance-to-airburst test results shown in Figure 1 again demonstrated that the mPE films can elongate more than the PVC films before failing, as previously shown by results of the physical property tests. The PVC bags burst along the weld edge after air pressure was applied for 72–105 seconds. The bags made with Type 1 mPE failed after 170–184 seconds, and those with Type 2 mPE film failed after 269–288 seconds—in each case because the body of the bag had been stretched beyond the yield point. The results for all films were not shown to be directly related to film thickness.

57b01.jpg

Figure 1. Resistance to airburst of a 432-cm2 bag filled with 12 psi air at 5.4 L/min.

57b02.jpg

Figure 2. Resistance to impact of a 432-cm2 bag with 620 ml water dropped on concrete. (*Denotes that film did not fail.)

The BSI standard for resistance to impact damage of a water-filled bag requires that it not leak or fail after being dropped from a height of 0.5 m (1.6 ft). None of the tested mPE or PVC films failed the 0.5-m drop, so the drop height was increased to determine the failure point. The PVC films failed along the weld edge at a drop height of 2.5–3.1 m (failure was not related to thickness), while none of the tested mPE films failed at the maximum drop height of 7.6 m (25 ft).

The BSI standard for resistance to bursting of a water-filled bag specifies that it must withstand a 350-N (78.7-lb) load without leaking or failing. None of the tested mPE or PVC films failed at a 350-N load, so the load was increased to determine a failure point. The maximum load that could be applied with the test apparatus was 2300 N (516 lb). Within this upper load limit, only one of the tested films failed: the mPE 1-6 film (0.15-mm thick), which failed at a load of 1410 N (316 lb). All of the other films—mPE and PVC—withstood the 2300-N load.

CONCLUSION
The films made with metallocene polyethylene resins that were used in this study showed superior physical and performance properties in several areas as compared with flexible PVC films. These included tensile strength, elongation, and toughness; resistance to puncture, impact, and bursting; and water-vapor barrier characteristics.

The study also demonstrated that mPE films can provide physical properties comparable to significantly thicker PVC films. For example, a 0.15-mm-thick mPE film can provide better toughness and puncture, impact, and burst resistance than a 0.25-mm-thick PVC film. This can enable product designers to specify a much thinner film—reducing thickness by 25% or more compared with PVC—without compromising performance.

The combination of a film density approximately 30% lower than that of PVC and improved properties results in a thinner, lighter-weight product that meets performance needs while reducing the volume of material required—making the product lighter to ship and use and creating a lower volume of waste material for disposable devices. The higher yield can also allow mPE films to be very cost-competitive: although PVC resins cost less per pound than mPE resins, the price per unit area of film or per finished device can be comparable.

In addition to the physical properties noted, mPE films offer a number of performance attributes that make them well suited for use in many medical devices. Among these properties are excellent flexibility over a wide temperature range, good cold crack and pinhole resistance, toughness and impact resistance at low temperatures, and the ability to be sterilized either by EtO or by gamma irradiation (with no discoloration)

Fabrication of devices with mPE films can be accomplished using many types of operations, including heat and impulse sealing, adhesive bonding and lamination, thermoforming, and printing. The mPE films can also be welded using conventional RF welding equipment modified with a catalyst film, with strong welds and tear seals achievable by this method. Solvent bonding—a technique commonly used for joining PVC components—does not work well with mPE films.

For manufacturers of a wide range of medical devices—from collection bags to drainage systems to inflatable devices—mPE films can provide a high-performance, practical, and cost-effective alternative to PVC films.

REFERENCES
1. Disposable Medical Supplies, Cleveland, The Fredonia Group, 1996.

. Shang S, and Woo L, "Selecting Materials for Medical Products: From PVC to Metallocene Polyolefins," Med Dev Diag Indust, 18(10):132–146, 1996

3. Hong KZ, "Poly(vinyl chloride) in Medical and Packaging Applications," in Society of Plastics Engineers, Inc., Technical Papers, vol XLI (ANTEC 95), Brookfield, CT, Society of Plastics Engineers, pp 4192–4198, 1995.

4. Moldovan D, "POPs and POEs—Resins for the 21st Century," in Proceedings from the Medical Design and Manufacturing Orlando Regional Show Series (supplement), Santa Monica, CA, Canon Communications, 1995.

5. Storer R (ed), Annual Book of ASTM Standards, West Conshohocken, PA, ASTM, 1993.

6. Wood L, and Keynes M, "BS 7126 Urine Collection Bags," London, British Standards Institution, 1989.