C101B Gas Permeability Test System is based on the differential pressure method,and is professionally applicable to the determination of gas transmission rate ofplastic films, composite films, high barrier materials, sheets, and metal foilsat different temperatures. The test system conforms to GB, ISO, ASTM and otherinternational standards.
Product FeaturesNote1
- Applicable to testing the permeability of various gases (exclude flammable, explosive and other hazardous gases)
- Support temperature control to satisfy tests under different conditions (optional)
- The processes of vacuumizing, pressure maintaining, start test and end test are automated.
- Automatic and manual test mode.
- Built-in high-quality stainless steel test cell with better sealing performance.
- Imported manual isolation valve is adopted for better sealing performance and lower failure rate
- Industrial computer appearance design, small size and fast cooling
- Reference film is provided for quick calibration
- The system adopts single-chip microcomputer control and can run independently
- The test report can be exported in common formats such as EXCEL and PDF
- Support micro printer, automatically print test data (optional)
- Equipped with RS232 data interface, which can be connected to computer software for curve analysis, data storage, report printing, etc. (optional)
Test Principle
The pre-conditioned specimen is mounted in the test cell as to form a sealed barrier between two chambers. The lower-pressure chamber (lower chamber) is firstly evacuated, followed by the evacuation of the entire cell. A flow of gas is thereafter introduced into the evacuated higher-pressure chamber (upper chamber) and a constant pressure (adjustable) difference is generated between the two chambers. The gas permeates through the specimen from higher pressure side into the lower pressure side. The gas permeability of the specimen can be obtained by analyzing the pressure changes in the lower chamber.
Standard
ISO 15105-1、ISO 2556、GB/T 1038-2000、ASTM D1434、JIS K7126-1、YBB 00082003
Applications
Applications |
---|
Films Gas transmission rate test of various plastic films, paper-plastic composite films, coextruded films, aluminized films, aluminum foils, aluminum foil composite films, glass fiber aluminum foil composite films and many others |
Sheets Gas transmission rate test of PP, PVC and PVDC sheets, metal foils, rubber pads, silicon wafers and other sheet materials |
Technical Specifications
Table 1: Test Parameters Note2
Parameters/Model | C101B |
---|
Testrange | cm3/m2·24h·0.1MPa | 0.1~5,000 |
Resolution | cm3/m2·24h·0.1MPa | 0.01 |
Test temperature | ℃ | 15~50(Optional) |
Temperature resolution | ℃ | 0.1 |
Temperature fluctuation | ℃ | ±0.5 |
Vacuum Resolution | Pa | 1 |
Vacuum Degree of Test Chamber | Pa | ≤ 20 |
Pressure difference | kPa | 101 |
Additional Functions | GMP Computer System requirement | Optional |
TC03 Temperature control device | Optional |
Table 2: Technical Specifications
Test Cell | 1 Cell |
---|
Specimen Size | 3.8” x 3.8”(9.7cm×9.7cm) |
---|
Specimen Thickness | ≤120 Mil(3mm) |
---|
Standard test area | 38.48cm2 |
---|
Test Gas | O2、N2、CO2 etc.(Outside of supply scope) |
---|
Gas Pressure | 72.5 PSI / 500 kPa |
---|
Port Size | Φ6 mm PU tubing |
---|
Instrument Dimension | 12.9” H x 16.9” W x 15.7” D (33cm× 43cm× 40cm) |
---|
Power Supply | 120VAC±10% 60Hz / 220VAC±10% 50Hz(one of two) |
---|
Net Weight | 50Lbs(23kg) |
---|
Table 3: Product Configuration
Standard Configuration | Instrument mainframe, vacuum pump (China), sampler, vacuum grease, Φ6 mm PU tubing |
---|
Optional Parts | Professional software, TC03 temperature control device, GMP Computer System requirement, Micro-printer |
---|
Note 1: The described product functions are subject to the specification in "Technical Parameters" "Table 1: Test Parameters".
Note 2: The parameters in the table are measured in Labthink laboratory by professional operators according to the requirements and conditions stapulated in laboratory environmental standards.
Contact Us!We can help you select or customize the proper unit for your test requirements.